In addition to regulatory pressure, industry watchers expect this situation to influence how health systems evaluate new procurement contracts. Some hospitals may pause new technology acquisitions until they see proof of corrective action implementation. The warning letter also demonstrates how important traceability, documentation discipline, and strong internal audit programs are in medical manufacturing. Clear communication between engineering teams, quality specialists, and regulatory units is critical to maintaining long-term trust. For a large global manufacturer like Philips, alignment between operational processes and compliance frameworks will now be monitored much more closely by regulators and buyers.
The U.S. Food and Drug Administration (FDA) has issued a warning letter to Philips, citing multiple quality system violations at three facilities responsible for manufacturing ultrasound equipment and cardiac imaging software.
The letter follows FDA inspections conducted earlier in 2025 at Philips sites in Washington, Pennsylvania, and the Netherlands. Inspectors identified shortcomings in the company’s procedures for managing product complaints, corrective actions, and device reporting.
Gaps in Complaint Handling and Documentation
According to the FDA, Philips’ complaint-handling unit failed to provide adequate documentation showing that reported issues were properly assessed and investigated. One example cited involved a 2023 report of an ultrasound probe that broke in half, where the investigation record was incomplete.
Another 2024 report described performance issues with a transesophageal ultrasound transducer used in three separate examinations. Despite clear indications of a device malfunction, Philips closed the file as a non-complaint, which the FDA said was inappropriate.
The agency also flagged instances where ultrasound transducers were refurbished and redistributed to customers after their three-year service life, contrary to recommended practice. Inspectors identified at least seven such devices, all linked to customer complaints.
Delayed Reporting of Device Corrections
The warning letter further noted that Philips failed to notify the FDA within 10 days of certain device corrections or removals, as required by regulation. One example involved a 2022 communication to customers about a voltage issue that could damage transducers over time. Although Philips subsequently corrected the software problem, the company did not submit a formal report of the corrective action to the agency.
Philips’ Response and Ongoing Quality Program
In a statement, Philips said it is implementing a multi-year program to strengthen its global quality management systems and resolve historical compliance issues. The company confirmed that nine of its facilities underwent FDA inspections earlier this year, with three resulting in documentation and process-related findings that are being addressed.